The use of the isovolume manoeuvre method as a calibration technique for respiratory monitoring instrumentation that detects the movement of the ribcage and the abdominal wall is analysed based on a model of two connected cylinders whose radii and heights may vary, and evidence is presented which suggests that this calibration method is robust in most circumstances. Some possible functional forms relating the variations in cylinder radius and height are examined, and methods for obtaining calibration constants based on these functional forms, purely from measurements of variations in the cylinder radius, are presented.