Key result
A multimodal wrist biosensor using the f14-PTT feature achieved high correlation with reference blood pressure and low root-mean-square error.
Why the study?
There is a need for a cuff-less, non-intrusive continuous blood pressure monitoring system utilizing a single measurement site for maximum wearability and patient convenience.
Does a multimodal wrist biosensor measuring PPG and IPG accurately estimate blood pressure in young healthy subjects?
Observational (n=10)
No
Does a multimodal wrist biosensor measuring PPG and IPG accurately estimate blood pressure in young healthy subjects?
Answers are grounded in this paper and cite their sources.
A novel multimodal wrist biosensor using PPG and IPG signals demonstrated adequate accuracy for cuff-less blood pressure estimation in a small cohort of healthy subjects.
We propose a multimodal biosensor for use in continuous blood pressure (BP) monitoring system. Our proposed novel configuration measures photo-plethysmography (PPG) and impedance plethysmography (IPG) signals simultaneously from the subject wrist. The proposed biosensor system enables a fully non-intrusive system that is cuff-less, also utilize a single measurement site for maximum wearability and convenience of the patients. The efficacy of the proposed technique was evaluated on 10 young healthy subjects. Experimental results indicate that the pulse transit time (PTT)-based features calculated from an IPG peak and PPG maximum second derivative (f 14 ) had a relatively high correlation coefficient ( r ) to the reference BP, with −0.81 ± 0.08 and −0.78 ± 0.09 for systolic BP (SBP) and diastolic BP (DBP), respectively. Moreover, here we proposed two BP estimation models that utilize six- and one-point calibration models. The six-point model is based on the PTT, whereas the one-point model is based on the combined PTT and radial impedance (Z). Thus, in both models, we observed an adequate root-mean-square-error estimation performance, with 4.20 ± 1.66 and 2.90 ± 0.90 for SBP and DBP, respectively, with the PTT BP model; and 6.86 ± 1.65 and 6.67 ± 1.75 for SBP and DBP, respectively, with the PTT-Z BP model. This study suggests the possibility of estimating a subject’s BP from only wrist bio-signals. Thus, the six- and one-point PTT-Z calibration models offer adequate performance for practical applications.
No takes yet. Share an insight, caveat, or question.
Rachim et al. (2019) conducted an observational in Healthy (n=10). Multimodal wrist biosensor (PPG and IPG) vs. Oscar 2 ambulatory blood pressure monitor was evaluated on Correlation coefficient (r) of f14-PTT feature to reference systolic and diastolic blood pressure (r = -0.81 for SBP, -0.78 for DBP). A multimodal wrist biosensor using the f14-PTT feature achieved high correlation with reference blood pressure and low root-mean-square error.