An 8-channel coil boosts carotid MRI vessel wall SNR ~1.7-fold over a 4-channel coil.
Answers are grounded in this paper and cite their sources.
PURPOSE: To design and compare an eight-channel phased array (PA) coil for carotid imaging to an established four-channel PA design at 3T. MATERIALS AND METHODS: An eight-channel PA (8PA) coil was designed specifically for imaging the carotid bifurcation and compared with the existing four-channel (4PA) design using a phantom and by in vivo black-blood magnetic resonance imaging (MRI). The 8PA and 4PA were compared in terms of coverage, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). RESULTS: The 8PA showed up to 1.7-fold improvement in SNR at a depth of 3.5 cm and greater longitudinal coverage at a given SNR on a phantom. The 8PA showed improved vessel wall SNR for high spatial resolution (0.63 mm(2)) PD, T1, and T2 (1.7, 1.7, 1.6 times, respectively; P <or= 0.002) and improved CNR (1.7, 1.6, 1.5 times, respectively; P <or= 0.002). Ultrahigh-resolution (0.27 mm(2)) T1-weighted images showed better SNR and CNR (1.4 times, P <or= 0.0001) on 8PA compared to 4PA. CONCLUSION: Carotid imaging studies may benefit from the improved SNR and larger coverage provided by use of the 8PA.
No takes yet. Share an insight, caveat, or question.
Balu et al. (2009) studied Carotid plaque imaging. Eight-channel phased array (8PA) coil at 3T vs. Four-channel phased array (4PA) coil at 3T was evaluated on Coverage, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) (1.6 to 1.7-fold SNR improvement; 1.5 to 1.7-fold CNR improvement, p=<=0.002). The 8-channel phased array coil improved carotid MRI vessel wall SNR and CNR compared to the 4-channel coil at 3T.
Synapse has enriched 10 related papers. Consider them for comparative context: