Answers are grounded in this paper and cite their sources.
The light output of thallium-activated cesium iodide was measured as a function of bombarding energy for protons (240 kev<~Eₚ<~2550 kev), alpha particles (980 kev<~E_α<~4920 kev), and C¹² ions (320 kev<~EC<~1850 kev). The relation of the light output to the particle energy is approximately linear in these energy ranges. The light outputs for different particles of the same energy are in the ratio p:α:C¹²=1.00:0.59:0.35. Saturation effects account for these results. The resolution varies inversely as the square root of the light output.
No takes yet. Share an insight, caveat, or question.
Bashkin et al. (1958) studied this question.