In this paper a respiratory mechanics model is considered, which is characterized by a biquadratic input impedance, and a sensitivity analysis has been carried out to determine the influence of experimental conditions on parameter estimation. This analysis was effected with data obtained experimentally, in three different patients under intermittent positive pressure ventilation. In all three cases, the model's input impedance demonstrated a maximum sensitivity in relation to the various parameters included in the field of frequencies from 0 to 10 Hz. This seems to suggest therefore, that the use of a low-pass filter with a cut-off frequency equal to 10 Hz could improve the signal/noise ratio and, consequently, the accuracy of the estimation of the parameters. Furthermore, the use of a system input with a bandwidth of 0-10 Hz provides the experimental conditions, under which good estimates of the parameters can be obtained. This conclusion has also been confirmed by simulation studies which have been conducted with different types of input signals.