Key result
SunTech Oscar2 records ~9-mmHg higher 24-hour systolic BP than comparator ambulatory monitors.
Why the study?
The U.S. Preventive Services Task Force recommends 24-hour ABPM for hypertension screening and diagnosis, but the optimal device for population-based surveys is unknown.
Do different 24-hour ABPM devices (Welch Allyn, Sun Tech, Spacelabs) differ in valid BP readings, mean BP values, and sleep experience in adults?
RCT (n=365)
Open-label
Stratified randomization (1:1:1 by sex, age group, and device type)
No
Do different 24-hour ABPM devices (Welch Allyn, Sun Tech, Spacelabs) differ in valid BP readings, mean BP values, and sleep experience in adults?
Different ABPM devices may yield significantly different systolic blood pressure readings despite similar rates of valid measurements and sleep interference.
Answers are grounded in this paper and cite their sources.
Device choice may alter systolic BP readings in ABPM; challenges interchangeability assumptions among commercial monitors.
BACKGROUND: The U.S. Preventive Services Task Force recommends the use of 24-hour ambulatory blood pressure monitoring (ABPM) as part of screening and diagnosis of hypertension. The optimal ABPM device for population-based surveys is unknown. METHODS: We compared the proportion of valid blood pressure (BP) readings, mean awake and asleep BP readings, differences between awake ABPM readings and initial standardized BP readings, and sleep experience among three ABPM devices. We randomized a convenience sample of 365 adults to 1 of 3 ABPM devices: Welch Allyn Mobil-O-Graph (WA), Sun Tech Classic Oscar2 (STO) and Spacelabs 90227 (SL). Participants completed sleep quality questionnaires on the nights before and during ABPM testing. RESULTS: The proportions of valid BP readings were not different among the 3 devices (P > 0.45). Mean awake and asleep systolic BP were significantly higher for STO device (WA vs. STO vs. SL: 126.65, 138.09, 127.44 mm Hg; 114.34, 120.34, 113.13 mm Hg; P < 0.0001 for both). The difference between the initial average standardized mercury systolic BP readings and the ABPM mean awake systolic BP was larger for STO device (WA vs. STO. vs. SL: -5.26, -16.24, -5.36 mm Hg; P < 0.0001); diastolic BP mean differences were ~ -6 mm Hg for all 3 devices (P = 0.6). Approximately 55% of participants reported that the devices interfered with sleep; however, there were no sleep differences across the devices (P > 0.4 for all). CONCLUSION: Most of the participants met the threshold of 70% valid readings over 24 hours. Sleep disturbance was common but did not interfere with completion of measurement in most of the participants.
No takes yet. Share an insight, caveat, or question.
Nwankwo et al. (2020) conducted an RCT in Hypertension screening / 24-hour ambulatory blood pressure monitoring (n=365). Sun Tech Classic Oscar2 (STO) and Welch Allyn Mobil-O-Graph (WA) 24-hour ABPM devices vs. Spacelabs 90227 (SL) 24-hour ABPM device (legacy reference) was evaluated on Mean 24-hour systolic blood pressure (mmHg) (MD 8.73 mmHg (STO vs SL), p=<0.0001). Sun Tech Oscar2 recorded significantly higher mean awake (138.09 vs 126.65 and 127.44 mmHg; P<0.0001) and asleep systolic BP than Welch Allyn and Spacelabs devices, with similar valid reading rates (P>0.45).