Key result
An automated patch clamp module recording from 384 cells simultaneously achieved success rates of >85% for completed experiments with various cell lines and 52% for stem cell-derived cardiomyocytes.
The development of a 384-channel automated patch clamp module enables high-throughput screening of ion channels with high success rates, including in stem cell-derived cardiomyocytes.
Answers are grounded in this paper and cite their sources.
May support scalable automated electrophysiology in research; leaves open validation for stem cell-derived cardiomyocyte applications.
We have developed an automated patch clamp module for high-throughput ion channel screening, recording from 384 cells simultaneously. The module is incorporated into a laboratory pipetting robot and uses a 384-channel pipettor head for application of cells and compounds. The module contains 384 amplifier channels for fully parallel recordings using a digital amplifier. Success rates for completed experiments (1- to 4-point concentration–response curves for cells satisfying defined quality control parameters) of greater than 85% have been routinely achieved with, for example, HEK, CHO, and RBL cell lines expressing hNa V 1.7, hERG, Kir2.1, GABA, or glutamate receptors. Pharmacology experiments are recorded and analyzed using specialized software, and the pharmacology of hNa V 1.7 and hERG is described. Fast external solution exchange rates of <50 ms are demonstrated using Kir2.1. Short exposure times are achieved by stacking the external solutions inside the pipette of the robot to minimize exposure of the ligand on the receptor. This ensures that ligand-gated ion channels, for example, GABA and glutamate described in this report, can be reproducibly recorded. Stem cell–derived cardiomyocytes have also been used with success rates of 52% for cells that have a seal resistance of >200 MΩ, and recordings of voltage-gated Na + and Ca 2+ are shown.
No takes yet. Share an insight, caveat, or question.
Obergrussberger et al. (2015) studied this question. Automated patch clamp module was evaluated on Success rates for completed experiments. An automated patch clamp module recording from 384 cells simultaneously achieved success rates of >85% for completed experiments with various cell lines and 52% for stem cell-derived cardiomyocytes.
Synapse has enriched 10 related papers. Consider them for comparative context: