Why the study?
Do small size (19-mm or 21-mm) ATS aortic valve prostheses provide satisfactory clinical outcomes compared to 25-mm valves in patients undergoing aortic valve replacement?
Do small size (19-mm or 21-mm) ATS aortic valve prostheses provide satisfactory clinical outcomes compared to 25-mm valves in patients undergoing aortic valve replacement?
Small (19-mm and 21-mm) ATS aortic valve prostheses perform satisfactorily in patients with a body surface area <1.6 m2, offering a viable alternative to annular enlargement despite higher transprosthetic gradients.
Answers are grounded in this paper and cite their sources.
Small ATS valves may be acceptable short-term in low-BSA patients; leaves open need for RCTs versus annular enlargement.
Records of 33 patients (group 1) who received 19-mm or 21-mm ATS aortic valve prostheses were retrospectively reviewed and compared with those of 26 patients who received 25-mm ATS valves (group 2). Group 1 patients were younger (20 ± 2 years versus 43 ± 15 years) and had a smaller mean body surface area (1.6 m 2 versus 1.83 m 2 ). Valve pathology was usually congenital aortic stenosis in group 1 and rheumatic in group 2. There was no operative mortality. The mean postoperative Doppler resting gradient was higher in group 1 (21 mm Hg versus 7 mm Hg). Total follow-up was 294 patient-years. There was no perivalvular leak, valve thrombosis, or endocarditis in either group. In group 2, there was 1 anticoagulant-related hemorrhage and 1 late death due to congestive heart failure. We concluded that in spite of the higher transprosthetic gradient, 19-mm and 21-mm ATS aortic valve prostheses performed satisfactorily in adult patients with a body surface area of less than 1.6 m 2 . This valve can be used in a small aortic annulus as an alternative to performing an annular enlargement procedure.
No takes yet. Share an insight, caveat, or question.
Demirkılıç et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: