A microbial biosensor system was based on an immobilized microorganism ( Succharomyces uvarum 4288 ATCC9080) and Clark‐type oxygen electrode consisting of a platinum cathode, lead anode, alkaline electrolyte, and an oxygen permeable Teflon membrane. The determination was based on the respiratory activity of the microorganism in presence of vitamin B 6 . Optimum conditions were: concentration of immobilized cells on membrane: OD 660 , = 0.05, temperature: 30°C. When oxygen permeable Teflon membranes (0.5 mil and 2.0 mil) were employed, relationships between vitamin B 6 , (pyridoxine) and current decrease were linear in the range of 0.5 ng/mL‐2.5 ng/mL and 2.5 ng/mL‐12.5 ng/mL, respectively. One assay could be completed within 15 min. This system provided rapid and simple determinations of vitamin B 6 in marine products with good correlation to traditional microbial assay values.
No takes yet. Share an insight, caveat, or question.
Endo et al. (1995) studied this question.