This is the authors' abstract. We don't add key points for this paper.
We describe a procedure for the simultaneous immunofluorometric assay of lutropin and follitropin in human serum, based on the use of monoclonal antibodies and of the fluorescent lanthanides Eu3+ and Tb3+. The alpha-chain-specific antibody was used as a common capture antibody on the surfaces of microtitration strips. The anti-beta-follitropin antibody was labeled with Tb3+, the anti-beta-lutropin antibody with Eu3+. After the immunoreactions had taken place, the bound fractions of the labels were dissociated in a fluorescence enhancement solution of pivaloyltrifluoroacetone, trioctylphosphine oxide, and Triton X-100 surfactant. In this solution both lanthanides can be measured successively with a time-resolved fluorometer. The detection limit of the assay is 0.1 int. unit/L for lutropin and 1 int. unit/L for follitropin. Results correlated well with those by commercial immunofluorometric assays and radioimmunoassays.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Hemmilä et al. (1987) studied this question.