We have proposed and demonstrated a novel method that can determine both the geometrical thickness and refractive index of a silicon wafer at the same time using an optical comb. The geometrical thickness and refractive index of a silicon wafer was determined from the optical thickness using phase information obtained in the spectral domain. In a feasibility test, the geometrical thickness and refractive index of a wafer were measured to be 334.85 microm and 3.50, respectively. The measurement uncertainty for the geometrical thickness was evaluated as 0.95 microm (k = 1) using a preliminary setup.
No takes yet. Share an insight, caveat, or question.
Jin et al. (2010) studied this question.