Inadequate connection between the S-ICD lead and header is a rare cause of electrical noise and inappropriate shocks that can be resolved by cleaning and reinserting the lead, avoiding unnecessary system extraction.
Implantable cardioverter-defibrillators (ICDs) effectively prevent deaths resulting from ventricular tachyarrhythmias.1,2 Because of complications and lead failures associated with transvenous ICD systems, an entirely extravascular subcutaneous ICD (S-ICD) was developed.3 A large, multicenter, randomized trial suggested that the S-ICD was noninferior to traditional transvenous ICD systems with respect to delivering appropriate therapy, avoiding inappropriate therapy, and avoiding device-related complications,4 although the interpretation of these data has been questioned.