Why the study?
Manual drawing of T1 maps for cardiomyopathies is time consuming and prone to subjective errors.
Does a deep learning algorithm provide automated measurements of native T1 and ECV comparable to manual radiologist analysis in patients with cardiomyopathies and healthy volunteers?
Does a deep learning algorithm provide automated measurements of native T1 and ECV comparable to manual radiologist analysis in patients with cardiomyopathies and healthy volunteers?
A commercially available deep learning algorithm can automatically and accurately measure native T1 and ECV fractions on CMR, achieving performance comparable to expert radiologists.
DL algorithm yields T1/ECV values comparable to radiologists; leaves open prospective validation before routine CMR adoption.
OBJECTIVE: T1 mapping provides valuable information regarding cardiomyopathies. Manual drawing is time consuming and prone to subjective errors. Therefore, this study aimed to test a DL algorithm for the automated measurement of native T1 and extracellular volume (ECV) fractions in cardiac magnetic resonance (CMR) imaging with a temporally separated dataset. MATERIALS AND METHODS: CMR images obtained for 95 participants (mean age ± standard deviation, 54.5 ± 15.2 years), including 36 left ventricular hypertrophy (12 hypertrophic cardiomyopathy, 12 Fabry disease, and 12 amyloidosis), 32 dilated cardiomyopathy, and 27 healthy volunteers, were included. A commercial deep learning (DL) algorithm based on 2D U-net (Myomics-T1 software, version 1.0.0) was used for the automated analysis of T1 maps. Four radiologists, as study readers, performed manual analysis. The reference standard was the consensus result of the manual analysis by two additional expert readers. The segmentation performance of the DL algorithm and the correlation and agreement between the automated measurement and the reference standard were assessed. Interobserver agreement among the four radiologists was analyzed. RESULTS: = 0.987 (95% CI, 0.980-0.991) and bias of 0.7% (95% LOA, -2.8%-4.2%) on per-subject basis. Agreements between DL and each of the four radiologists were excellent (intraclass correlation coefficient [ICC] of 0.98-0.99 for both native T1 and ECV), comparable to the pairwise agreement between the radiologists (ICC of 0.97-1.00 and 0.99-1.00 for native T1 and ECV, respectively). CONCLUSION: The DL algorithm allowed automated T1 and ECV measurements comparable to those of radiologists.
No takes yet. Share an insight, caveat, or question.
Chang et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: