Why the study?
Minimal evidence was available regarding the reduction in healthcare utilization with remote care in ICD patients over longer follow-up.
Does remote care reduce healthcare utilization in patients with an implantable cardioverter-defibrillator or CRT-D?
Does remote care reduce healthcare utilization in patients with an implantable cardioverter-defibrillator or CRT-D?
Remote monitoring of ICD and CRT-D patients significantly reduces overall healthcare utilization over a nearly 3-year follow-up period compared to conventional in-office care.
Answers are grounded in this paper and cite their sources.
Remote care may lower long-term healthcare utilization in ICD patients; leaves open confirmation by randomized trials.
BACKGROUND: Minimal evidence is available of the reduction in healthcare utilization of remote care in ICD patients over a longer period of follow-up. OBJECTIVES: This study compared healthcare utilization up to 3 year follow-up in implantable cardioverter-defibrillator (ICD) patients with remote care compared to conventional care. METHODS: We conducted a retrospective cohort study of patients who received a single or dual-chamber ICD or cardiac resynchronization therapy-defibrillator (CRT-D) between 2016 and 2018. Patients with remote care and patients were compared with patients with received conventional care (control group). The primary endpoint was a composite of cardiac follow-up visits, ICD follow-up visits, telephone consultations, emergency department (ED) visits and hospital admissions and was defined as total healthcare utilization. The secondary endpoints were the individual care activities and one-year all-cause mortality. RESULTS: A total of 497 patients were included in the study, of which 299 patients were allocated to the remote care and 198 patients to the control group. Mean follow-up was 815 ± 279 days. Remote care was associated with a significantly lower rate of adjusted total healthcare utilization in comparison to the control group that sustained for 3 subsequent follow-up years (IRR = 0.78, 95% CI [0.67 to 0.92], p < .01). One-year all-cause mortality was similar between the remote care and control group (respectively 3.0% vs. 5.5%, p = .29). CONCLUSIONS: Compared to the standard follow-up of in-office care, a remote care program was associated with a sustained lower rate of planned and unplanned healthcare utilization up to 3 subsequent years after ICD/CRT-D implantation.
No takes yet. Share an insight, caveat, or question.
Steenbergen et al. (2021) studied this question.