Answers are grounded in this paper and cite their sources.
A digital chest radiography system has been developed, with a detector based on the photoelectric properties of amorphous selenium. The selenium layer is deposited on a cylindrical aluminium drum, large enough to cover the full field of view for chest imaging. The electrostatic charge image which is formed on the selenium surface after x-ray exposure is read out by electrometer probes using fast drum rotation. For a physical evaluation of the attainable image quality, the characteristic curve, the modulation transfer function, and the noise spectra were measured. From these measurements, the signal-to-noise properties of the detector in terms of detective quantum efficiency (DQE) and noise equivalent quanta (NEQ) were derived. The results show that the selenium-based detector has a wide dynamic range and a significantly better DQE than screen-film and storage phosphor systems for spatial frequencies below the Nyquist limit (2.7 lp/mm). As a consequence, the detectability of small, low-contrast details is considerably improved.
No takes yet. Share an insight, caveat, or question.
Neitzel et al. (1994) studied this question.