Answers are grounded in this paper and cite their sources.
The present work describes a validated reverse phase high performance liquid chromatographic method for simultaneous estimation of telmisartan and hydrochlorothiazide in tablet formulation. Chromatography was performed on a ODS Hypersil C18 (25 cm×4.6 mm I.D) column from thermo in isocratic mode with mobile phase containing acetonitrile:0.05 M KH2PO4 pH 3.0 (60:40). The flow rate was 1.0 ml/min and the eluent was monitored at 271 nm. The selected chromatographic conditions were found to effectively separate telmisartan (RT- 5.19 min) and hydrochlorothiazide (RT- 2.97 min). Linearity for telmisartan and hydrochlorothiazide were found in the range of 4.1-20.48 µg/ml and 1.28-6.4 µg/ml, respectively. The proposed method was found to be accurate, precise, reproducible and specific and can be used for simultaneous analysis of these drugs in tablet formulation.
No takes yet. Share an insight, caveat, or question.
Wankhede et al. (2007) studied this question.