Answers are grounded in this paper and cite their sources.
Evaluates a new method to characterize the size-of-source effect in thermal imaging cameras, suggesting improvements in measurement efficiency.
The size-of-source effect (SSE) is a major source of uncertainty in radiation thermometers and thermal imaging cameras. This effect is considered a systematic error and is typically evaluated by measuring changes in the detected signal as the size of the radiant source is varied using a set of circular apertures of different diameters. The accurate characterization of the SSE requires measurements over a wide range of object sizes with well-distributed sample points, which is both time-consuming and labor-intensive. This paper proposes a new approach in which the aperture of an iris diaphragm is continuously adjusted to vary the size of the radiant source. An experimental study was conducted to compare the results obtained with individual fixed apertures and to assess whether the camera frame rate and the iris-driving period influence the measured temperature. The proposed approach reproduced the SSE with the same accuracy as measurements made with individual apertures while providing a significantly larger number of measurement points and a notable reduction in the measurement time. Furthermore, the method showed no significant sensitivity to the driving speed of the iris or the camera's frame rate within the investigated range. However, temporal effects arising from the optical components were found to disturb the measurement.
No takes yet. Share an insight, caveat, or question.
Méndez-Bohórquez et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: