Answers are grounded in this paper and cite their sources.
PURPOSE: To compare the accuracy of a newly developed intraocular lens (IOL) power formula (VRF-G) with twelve existing formulas (Barret Universal II, EVO 2.0, Haigis, Hill-RBF 2.0, Hoffer Q, Holladay 1, Kane, Næeser 2, PEARL-DGS, SRK/T, T2 and VRF). METHODS: Retrospective case series including 828 patients having uncomplicated cataract surgery with the implantation of a single IOL model (SN60WF). Using optimised constants, refraction prediction error of each formula was calculated for each eye. Subgroup analysis was performed based on the axial length (short ≤22.0mm; medium >22.0mm to <26.0mm; long ≥26.0mm). Main outcomes included mean prediction error (ME) mean (MAE) and median absolute error (MedAE), in diopters (D), and the percentage of eyes within ±0.25D, ±0.50D, ±0.75D and ±1.00D. RESULTS: Formulas absolute errors were statistically different among them (p<0.001), with Kane having the lowest MAE of all formulas, followed by EVO 2.0 and VRF-G, which had the lowest MedAE. The Kane formula had the highest percentage of eyes within ±0.25D (47.0%) and ±1.00D (97.7%) and the VRF-G formula had the highest percentage of eyes within ±0.50D (79.5%). For all AL subgroups, Kane, EVO 2.0 and VRF-G formulas had the most accurate performances (lowest MAE). CONCLUSION: New generation formulas may help us in achieving better refractive results, lowering the variance in accuracy in extreme eyes - Kane, EVO 2.0 and VRF-G formulas are promising candidates to fulfil that goal.
No takes yet. Share an insight, caveat, or question.
Hipólito‐Fernandes et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: