STUDY DESIGN: A carbon fiber-reinforced polymer implant, designed to aid interbody lumbar fusion, was tested biologically in an experimental surgical model. Twenty-seven Spanish goats had interbody lumbar fusion surgery in a randomized protocol. Seventeen goats were implanted with the carbon fiber-reinforced polymer cage packed with autologous bone, and 10 goats were implanted with ethylene oxide-sterilized allograft bone. OBJECTIVES: To determine fusion success, biocompatibility of the carbon polymer material, and possibility of carbon wear debris at intervals after surgical implantation. METHODS: Goats were killed at 6 months, 12 months, and 24 months and full-body autopsies were done. Spine specimens were studied by plain radiography, three-dimensional reformatted computed tomography studies, and histology. RESULTS: At 6 months, one of three allograft implantations showed histologic and radiographic fusion, whereas five of five carbon fiber-reinforced polymer cage fusions showed at least partial fusion. At 12 months, two of three allograft implantations and five of five carbon fiber-reinforced polymer cage fusions were solidly fused. At 24 months, five of five allograft implantations and three of three carbon fiber-reinforced polymer cage implantations were solidly fused. CONCLUSIONS: Interbody fusion using a carbon cage implant packed with autologous bone achieved a quicker and more reliable fusion compared with ethylene oxide-sterilized allograft bone. There were no adverse effects from the implant material.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Brantigan et al. (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: