Key result
HPLC lipoprotein analysis demonstrates high accuracy for total cholesterol but overestimates HDL-C by ~11%.
Answers are grounded in this paper and cite their sources.
BACKGROUND: The main purpose of this study was to evaluate the between-instrument variation of the HPLC method for the measurement of total cholesterol (TC), HDL-cholesterol (HDL-C), LDL-cholesterol (LDL-C), VLDL-cholesterol (VLDL-C), chylomicron cholesterol (CM-C), LDL size, and HDL size. Furthermore, the accuracy of the HPLC was assessed for the determination of TC and HDL-C, compared with CDC reference methods. METHODS: We used four HPLC instruments with different column-load numbers from 250 to 5000. For accuracy assessment of TC and HDL-C, we used the reference methods recommended by the CDC. RESULTS: The values measured by the four instruments were highly correlated with each other (mean r = 0.965), and the absolute mean differences were 4-43 mg/L for TC, 4-30 mg/L for HDL-C, 0-48 mg/L for LDL-C, 7-66 mg/L for VLDL-C, 0-7 mg/L for CM-C, 0.1-0.3 nm for LDL size, and 0-0.1 nm for HDL size. For TC, the HPLC instruments showed high correlation and good agreement with the reference method: r = 0.997; total error <6.6%; absolute mean bias <1.2%. For HDL-C, the results from the HPLC method were significantly higher (10.8% absolute mean bias) than those of the CDC reference method, in spite of good correlation between the two methods (r = 0.998). CONCLUSIONS: The between-instrument variation in serum lipoprotein analysis by HPLC was confirmed to be very small. This method met the US National Cholesterol Education Program's performance criteria for TC but not for HDL-C.
No takes yet. Share an insight, caveat, or question.
Usui et al. (2000) studied Serum lipoprotein measurement. HPLC method for serum lipoproteins vs. CDC reference methods for TC and HDL-C was evaluated on Between-instrument variation and accuracy of HPLC method for serum lipoproteins compared with CDC reference methods (r = 0.965 (between-instrument); r = 0.997 (TC accuracy); r = 0.998 (HDL-C correlation)). HPLC serum lipoprotein analysis showed low between-instrument variation (mean r=0.965) and met accuracy criteria for TC (bias <1.2%, r=0.997) but over-estimated HDL-C (10.8% bias) vs CDC reference.
Synapse has enriched 5 related papers. Consider them for comparative context: