The Analytab Products Inc. (API), Micro-Drop (MD), and Uni-Yeast-Tek (UYT) systems for the presumptive identification of common clinical yeast isolates were compared with the oxidation-fermentation (OF) and a conventional procedure. With 229 coded isolates, the identification accuracies were API 94, MD 83, OF 82, and UYT 99%. The API system required the greatest technical ability. The MD materials were prone to malfunction. OF media, if incubated beyond 14 days, gave an accuracy of 87%, but this offered no advantage over the conventional procedure. The UYT system was the easiest to use.
No takes yet. Share an insight, caveat, or question.
Bowman et al. (1976) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: