A rapid and simple high performance liquid chromatography (HPLC) method with a UV detection (241 nm) was developed and validated for estimation of eplerenone from spiked human plasma. The analyte and the internal standard (valdecoxib) were extracted with a mixture of dichloromethane and diethyl ether. The chromatographic separation was performed on a HiQSil C-18HS column (250 mm×4.6 mm, 5 μm) with a mobile phase consisting of acetonitrile:water (50:50, v/v) at flow rate of 1 mL/min. The calibration curve was linear in the range 100–3200 ng/mL and the heteroscedasticity was minimized by using weighted least squares regression with weighting factor 1/X.
No takes yet. Share an insight, caveat, or question.
Gide et al. (2012) studied this question.