Compares discrepancies in oxygen consumption data recorded by EMR and automated monitoring systems, indicating integration benefits.
Accurate monitoring of oxygen therapy is essential for patient safety, resource management, and financial accountability. However, the high demands of the clinical environment may limit real-time documentation, potentially affecting the accuracy of manual electronic medical record (EMR) entries. Automated monitoring technologies, such as ATAS O₂ (Salvus, Pernambuco, Brazil), have emerged as a potential approach to address such limitations. This study compared oxygen consumption data recorded in the EMR with measurements obtained using ATAS O₂ across 21 hospital beds over a five-month period. The ATAS O₂ system detected a higher number of beds receiving oxygen, total duration (566.90 ± 372.56 vs. 163.61 ± 209.85 h, p < 0.001), number of cycles (138.14 ± 118.65 vs. 12.24 ± 10.00, p < 0.001), mean flow (2.70 ± 1.96 vs. 0.82 ± 0.77 L/min, p < 0.001), total volume (78.41 ± 44.84 vs. 4.72 ± 5.33 m³, p < 0.001), and volume per cycle (1,275.56 ± 2,052.22 vs. 385.78 ± 357.89 L, p = 0.010). These findings indicate systematic discrepancies between manual EMR and automated monitoring systems, highlighting the potential value of integrating automated monitoring technologies into hospital infrastructures.