Answers are grounded in this paper and cite their sources.
We report on methodology for employing a conventional linear transducer array as a thermoacoustic detector in a thermoacoustic computed tomography (TCT) device, which has been designed for imaging small animals, e.g., athymic nude mice. We tested this concept using a 5 MHz, 128-element linear array (Acuson model L538). Thermoacoustic emissions were induced in a tissue-mimicking phantom using a Nd:YAg laser, operated at 1064 nm. Two-dimensional, axial "slice" images were formed using a filtered-backprojection algorithm. In-plane spatial resolution was measured as better than 200 microns with a slice thickness of 1.5 mm (full width at half maximum). The same detector, when operated as a conventional phased array, produced conventional ultrasound images in perfect registration with the TCT images.
No takes yet. Share an insight, caveat, or question.
Kruger et al. (2003) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: