Answers are grounded in this paper and cite their sources.
• Polymer-brush-gated OECT enables enzyme-free, label-free glucose detection. • Solvent-controlled tubular poly(EDOT-FPBA) improves ion transport and sensitivity. • Sensor shows high selectivity and long-term stability in biological environments. • Wide detection range (0.001–40 mM) covers clinically relevant glucose levels. Non-enzymatic glucose detection remains challenging due to poor selectivity and instability in complex biological environments. Here, we developed an organic electrochemical transistor (OECT) incorporating a CH 2 Cl 2 -derived poly(EDOT-FPBA) gate to enhance glucose sensing performance. A solvent-controlled electropolymerization strategy was employed to construct an ordered polymer-brush architecture, enabling efficient ion transport and improved exposure of boronic-acid recognition sites. Structural and electrical characterizations confirmed the brush-like morphology and its strong gate modulation capability. The device exhibited a pronounced gate-voltage shift(ΔV G ) of ≈82.0 mV, a broad linear range (0.001–40 mmol⸱L -1 ), and a low detection limit (5.6 µmol⸱L -1 ), with interference below 15%, repeatability within a relative standard deviation of 4.6%, and serum recoveries of 92.0–107.4%. This strategy provides a promising route for developing low-voltage, enzyme-free OECT for reliable glucose monitoring in complex biological systems.
No takes yet. Share an insight, caveat, or question.
Bai et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: