Key result
Intra-aortic balloon pump 1:1 counterpulsation exacerbated the peak LVOT gradient to 114 mm Hg, whereas placing the device on standby reduced the gradient to <20 mm Hg and increased MAP by 15 mm Hg.
Why the study?
Does intra-aortic balloon pump counterpulsation worsen left ventricular outflow tract obstruction compared to standby mode in a patient with combined hypertrophic and takotsubo cardiomyopathy in cardiogenic shock?
Case Report (n=1)
No
Does intra-aortic balloon pump counterpulsation worsen left ventricular outflow tract obstruction compared to standby mode in a patient with combined hypertrophic and takotsubo cardiomyopathy in cardiogenic shock?
In patients with combined hypertrophic and takotsubo cardiomyopathy complicated by cardiogenic shock, IABP counterpulsation and inotropes can severely exacerbate dynamic LVOT obstruction and should be avoided in favor of pure alpha-agonists.
Answers are grounded in this paper and cite their sources.
IABP counterpulsation may worsen LVOT obstruction in hypertrophic cardiomyopathy shock; leaves open optimal support strategies in this rare presentation.
•Obstructive HCM with superimposed takotsubo syndrome led to shock.•The use of an IABP worsened outflow obstruction.•Putting the IABP on standby improved outflow tract gradients dramatically.•An IABP should not be used in shock with LVOT obstruction.
No takes yet. Share an insight, caveat, or question.
Griffin et al. (2023) conducted a case report in Hypertrophic cardiomyopathy complicated by takotsubo cardiomyopathy and cardiogenic shock (n=1). Intra-aortic balloon pump (IABP) 1:1 counterpulsation vs. Intra-aortic balloon pump (IABP) on standby / removal and phenylephrine was evaluated on Peak left ventricular outflow tract (LVOT) gradient on continuous-wave spectral Doppler echocardiography. Intra-aortic balloon pump 1:1 counterpulsation exacerbated the peak LVOT gradient to 114 mm Hg, whereas placing the device on standby reduced the gradient to <20 mm Hg and increased MAP by 15 mm Hg.