Key result
A noise-reducing fabric electrode with a folded structure improved SNR values by about 12.02 dB at rest and 10.29 dB during exercise compared to a standard fabric electrode.
Why the study?
Fabric electrodes used for long-term ECG measurements introduce noise due to large skin-electrode impedance, conventionally requiring glued chip-type passive elements.
Does a noise-reducing fabric electrode improve SNR in ECG measurement compared to a standard fabric electrode?
Does a noise-reducing fabric electrode improve SNR in ECG measurement compared to a standard fabric electrode?
A novel folded fabric electrode structure effectively acts as a low-pass filter, significantly improving the signal-to-noise ratio during ECG measurements at rest and during exercise.
May reduce noise in long-term fabric ECG; leaves open clinical validation before practice change.
In this work, we propose a fabric electrode with a special structure that can play the role of a noise reduction filter. Fabric electrodes made of the conductive fabric have been used for long-term ECG measurements because of their flexibility and non-invasiveness; however, due to the large impedance between the skin and the fabric electrodes, noise is easily introduced into the ECG signal. In contrast to conventional work, in which chip-type passive elements are glued to the electrode to reduce noise, the proposed electrode can obtain a noise-reduced ECG by changing the structure of fabric. Specifically, the proposed electrode was folded multiple times to form a capacitor with a capacitance of about 3 nF. It is combined with the skin-electrode impedance to form a low-pass filter. In the experiment, we made a prototype of the electrodes and measured ECG at rest and during EMG-induced exercise. As a result, the SNR values at rest and during exercise were improved about 12.02 and 10.29 dB, respectively, compared with the fabric electrode without special structure. In conclusion, we have shown that changing the fabric electrode structure effectively removes noise in ECG measurement.
No takes yet. Share an insight, caveat, or question.
Terada et al. (2021) studied ECG measurement. Noise-reducing fabric electrode with folded structure vs. Fabric electrode without special structure was evaluated on SNR values at rest and during EMG-induced exercise (improved about 12.02 and 10.29 dB). A noise-reducing fabric electrode with a folded structure improved SNR values by about 12.02 dB at rest and 10.29 dB during exercise compared to a standard fabric electrode.