Why the study?
The study was conducted to characterise the noise properties of un-gelled, textile-based electrodes intended for ECG recording in wearable garments.
This study characterizes the noise properties of textile-based dry ECG electrodes, providing baseline technical specifications for their use in wearable garments.
Textile electrode noise characterisation informs wearable ECG amplifier design; leaves open clinical validation in ambulatory use.
This study reports attempts to characterise the noise properties of un‐gelled, textile‐based electrodes intended for use in ECG recording in wearable garments. Measurements were made using very low noise operational amplifiers and with mains power supply interference eliminated from the recorded signals so that they reflected only genuine noise. Measurements were carried out on four conductive fabric electrodes, one conductive rubber electrode and a standard self‐adhesive, gelled electrode on three male and three female subjects. Curves of the form y = K / f + C were fitted to the recorded noise spectral density functions. Coefficients obtained for the white noise spectral density ranged from 1.1 to 5.6 nV/√Hz and for the flicker (1/ f ) noise spectral density ranged from 8 to 80 nV√Hz.