Key result
Modifying the ATS mechanical heart valve design by optimizing the leaflet-housing gap, increasing the opening angle, and introducing a streamlined channel improved its thrombogenic performance in vitro.
The Device Thrombogenicity Emulator methodology can successfully guide design modifications in mechanical heart valves to reduce thrombogenicity.
Answers are grounded in this paper and cite their sources.
Optimized valve designs reduce in vitro thrombogenicity; hypothesis-generating for emulator-guided modifications pending clinical validation.
Patients implanted with mechanical heart valves (MHV) or with ventricular assist devices that use MHV require mandatory lifelong anticoagulation for secondary stroke prevention. We recently developed a novel Device Thrombogenicity Emulator (DTE) methodology that interfaces numerical and experimental approaches to optimize the thrombogenic performance of the device and reduce the bleeding risk associated with anticoagulation therapy. Device Thrombogenicity Emulator uses stress-loading waveforms in pertinent platelet flow trajectories that are extracted from highly resolved numerical simulations and emulates these flow conditions in a programmable hemodynamic shearing device (HSD) by which platelet activity is measured. We have previously compared two MHV, ATS and the St. Jude Medical, and demonstrated that owing to its nonrecessed hinge design, the ATS valve offers improved thrombogenic performance. In this study, we further optimize the ATS valve thrombogenic performance, by modifying various design features of the valve, intended to achieve reduced thrombogenicity: 1) optimizing the leaflet-housing gap clearance; 2) increasing the effective maximum opening angle of the valve; and 3) introducing a streamlined channel between the leaflet stops of the valve that increases the effective flow area. We have demonstrated that the DTE optimization methodology can be used as test bed for developing devices with significantly improved thombogenic performance.
No takes yet. Share an insight, caveat, or question.
Alemu et al. (2010) studied Mechanical heart valve thrombogenicity. Design modifications of the ATS mechanical heart valve vs. Original ATS valve design was evaluated on Thrombogenic performance (platelet activity). Modifying the ATS mechanical heart valve design by optimizing the leaflet-housing gap, increasing the opening angle, and introducing a streamlined channel improved its thrombogenic performance in vitro.
Synapse has enriched 10 related papers. Consider them for comparative context: