Key result
CRT-D heart sound metrics correlate closely with invasive contractility and filling pressures in animal models.
Why the study?
Do heart sound parameters (S1, S3, and systolic time intervals) measured by an embedded accelerometer in an implanted CRT-D device correlate with acute invasive haemodynamic changes in contractility (LV dP/dtmax) and filling pressure (LAP) in animal models of ischaemia and pulmonary oedema?
Do heart sound parameters (S1, S3, and systolic time intervals) measured by an embedded accelerometer in an implanted CRT-D device correlate with acute invasive haemodynamic changes in contractility (LV dP/dtmax) and filling pressure (LAP) in animal models of ischaemia and pulmonary oedema?
Heart sound amplitudes and systolic time intervals measured via an existing CRT-D accelerometer track acute changes in cardiac contractility and left atrial filling pressure in preclinical models, supporting non-invasive multi-sensor remote heart failure monitoring without dedicated pressure sensor implants.
Answers are grounded in this paper and cite their sources.
Supports accelerometer-based heart sound monitoring in CRT-D devices for acute hemodynamics; leaves open clinical translation in human heart failure.
Aim The aim of this study was to evaluate the haemodynamic correlates of heart sound (HS) parameters such as third HS (S3), first HS (S1), and HS-based systolic time intervals (HSTIs) from an implantable cardiac device. Methods and results Two unique animal models (10 swine with myocardial ischaemia and 11 canines with pulmonary oedema) were used to evaluate haemodynamic correlates of S1, S3, and HSTIs, namely, HS-based pre-ejection period (HSPEP), HS-based ejection time (HSET), and the ratio HSPEP/HSET during acute haemodynamic perturbations. The HS was measured using implanted cardiac resynchronization therapy defibrillator devices simultaneously with haemodynamic references such as left atrial (LA) pressure and left ventricular (LV) pressure. In the ischaemia model, S1 amplitude (r = 0.76 ± 0.038; P = 0.002), HSPEP (r = −0.56 ± 0.07; P = 0.002), and HSPEP/HSET (r = −0.42 ± 0.1; P = 0.002) were significantly correlated with LV dP/dtmax. In contrast, HSET was poorly correlated with LV dP/dtmax (r = 0.14 ± 0.14; P = 0.23). In the oedema model, a physiological delayed response was observed in S3 amplitude after acute haemodynamic perturbations. After adjusting for the delay, S3 amplitude significantly correlated with LA pressure in individual animals (r = 0.71 ± 0.07; max: 0.92; min: 0.17) as well as in aggregate (r = 0.62; P < 0.001). The S3 amplitude was able to detect elevated LA pressure, defined as >25 mmHg, with a sensitivity = 58% and specificity = 90%. Conclusions The HS parameters such as S1, S3, and HSTIs measured using implantable devices significantly correlated with haemodynamic changes in acute animal models, suggesting potential utility for remote heart failure patient monitoring.
No takes yet. Share an insight, caveat, or question.
Thakur et al. (2017) studied Acute myocardial ischaemia and acute pulmonary oedema (animal models of heart failure) (n=21). Implantable CRT-D device with embedded accelerometer for heart sound (S1, S3, HSTI) monitoring vs. Invasive haemodynamic pressure catheter measurements (LV pressure dP/dtmax and left atrial pressure) was evaluated on Correlation between device-measured heart sound parameters (S1 amplitude, HSPEP, HSPEP/HSET, S3 amplitude) and invasive haemodynamic parameters (LV dP/dtmax and left atrial pressure) (r = 0.76 ± 0.038 (S1 vs LV dP/dtmax); r = 0.62 (S3 vs LAP aggregate), p=0.002 (S1 vs LV dP/dtmax); <0.001 (S3 vs LAP aggregate)). Implantable CRT-D heart sound metrics correlated with LV dP/dtmax and left atrial pressure in acute animal models.