Key result
Heart sound parameters measured using implantable devices significantly correlated with haemodynamic changes, including S3 amplitude with left atrial pressure (r=0.62, P<0.001).
Why the study?
Do heart sound parameters measured by an implanted cardiac device correlate with haemodynamic changes in animal models of ischaemia and pulmonary oedema?
Do heart sound parameters measured by an implanted cardiac device correlate with haemodynamic changes in animal models of ischaemia and pulmonary oedema?
Effect estimate: r = 0.62
p-value: p=<0.001
Heart sound parameters measured by implantable devices correlate significantly with acute haemodynamic changes in animal models, suggesting potential utility for remote heart failure monitoring.
Device HS parameters track haemodynamics in animal ischaemia/oedema models; hypothesis-generating for clinical implantable monitoring.
Aim The aim of this study was to evaluate the haemodynamic correlates of heart sound (HS) parameters such as third HS (S3), first HS (S1), and HS-based systolic time intervals (HSTIs) from an implantable cardiac device. Methods and results Two unique animal models (10 swine with myocardial ischaemia and 11 canines with pulmonary oedema) were used to evaluate haemodynamic correlates of S1, S3, and HSTIs, namely, HS-based pre-ejection period (HSPEP), HS-based ejection time (HSET), and the ratio HSPEP/HSET during acute haemodynamic perturbations. The HS was measured using implanted cardiac resynchronization therapy defibrillator devices simultaneously with haemodynamic references such as left atrial (LA) pressure and left ventricular (LV) pressure. In the ischaemia model, S1 amplitude (r = 0.76 ± 0.038; P = 0.002), HSPEP (r = −0.56 ± 0.07; P = 0.002), and HSPEP/HSET (r = −0.42 ± 0.1; P = 0.002) were significantly correlated with LV dP/dtmax. In contrast, HSET was poorly correlated with LV dP/dtmax (r = 0.14 ± 0.14; P = 0.23). In the oedema model, a physiological delayed response was observed in S3 amplitude after acute haemodynamic perturbations. After adjusting for the delay, S3 amplitude significantly correlated with LA pressure in individual animals (r = 0.71 ± 0.07; max: 0.92; min: 0.17) as well as in aggregate (r = 0.62; P < 0.001). The S3 amplitude was able to detect elevated LA pressure, defined as >25 mmHg, with a sensitivity = 58% and specificity = 90%. Conclusions The HS parameters such as S1, S3, and HSTIs measured using implantable devices significantly correlated with haemodynamic changes in acute animal models, suggesting potential utility for remote heart failure patient monitoring.
No takes yet. Share an insight, caveat, or question.
Thakur et al. (2017) studied Myocardial ischaemia and pulmonary oedema (n=21). Implantable cardiac device heart sound monitoring vs. Invasive haemodynamic references (LA and LV pressure) was evaluated on Correlation of S3 amplitude with left atrial pressure in the oedema model (r = 0.62, p=<0.001). Heart sound parameters measured using implantable devices significantly correlated with haemodynamic changes, including S3 amplitude with left atrial pressure (r=0.62, P<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: