Answers are grounded in this paper and cite their sources.
Criteria for terminating smoothing to remove a cubic spline background from the x-ray absorption coefficient are stated in terms of three parameters obtained from the k3 weighted Fourier transform of the resulting x-ray absorption fine structure data. The three parameters are HR, the average value of the transform magnitude between 0 and 0.25 Å, HM, the maximum value in the transform magnitude between 1 and 5 Å, and HN, the average value of the transform magnitude between 9 and 10 Å. The termination criteria are HR−HN ⩾0.05HM; or if HN ≳0.1HM, then HR ⩾0.1HM. The incorporation of the criteria into a computer program to facilitate automatic background removal is discussed. Examples of application of the technique to copper, β-PtO2, and ferritin samples are presented.
No takes yet. Share an insight, caveat, or question.
Cook et al. (1981) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: