Key result
Textile silver concentric ring electrodes provided a more stable response than PEDOT:PSS electrodes and higher spatial resolution than conventional ECG for bipolar concentric ECG recordings.
Why the study?
Do textile silver concentric ring electrodes provide better ECG signal quality and stability compared to PEDOT:PSS electrodes and conventional ECG?
Do textile silver concentric ring electrodes provide better ECG signal quality and stability compared to PEDOT:PSS electrodes and conventional ECG?
Textile silver concentric ring electrodes offer improved spatial resolution and stability for ECG recording compared to PEDOT:PSS electrodes and conventional ECG, bringing them closer to clinical use.
Textile CREs may enable comfortable high-resolution ECG; leaves open clinical validation before practice change.
Among many of the electrode designs used in electrocardiography (ECG), concentric ring electrodes (CREs) are one of the most promising due to their enhanced spatial resolution. Their development has undergone a great push due to their use in recent years; however, they are not yet widely used in clinical practice. CRE implementation in textiles will lead to a low cost, flexible, comfortable, and robust electrode capable of detecting high spatial resolution ECG signals. A textile CRE set has been designed and developed using screen-printing technology. This is a mature technology in the textile industry and, therefore, does not require heavy investments. Inks employed as conductive elements have been silver and a conducting polymer (poly (3,4-ethylenedioxythiophene) polystyrene sulfonate; PEDOT:PSS). Conducting polymers have biocompatibility advantages, they can be used with flexible substrates, and they are available for several printing technologies. CREs implemented with both inks have been compared by analyzing their electric features and their performance in detecting ECG signals. The results reveal that silver CREs present a higher average thickness and slightly lower skin-electrode impedance than PEDOT:PSS CREs. As for ECG recordings with subjects at rest, both CREs allowed the uptake of bipolar concentric ECG signals (BC-ECG) with signal-to-noise ratios similar to that of conventional ECG recordings. Regarding the saturation and alterations of ECGs captured with textile CREs caused by intentional subject movements, silver CREs presented a more stable response (fewer saturations and alterations) than those of PEDOT:PSS. Moreover, BC-ECG signals provided higher spatial resolution compared to conventional ECG. This improved spatial resolution was manifested in the identification of P1 and P2 waves of atrial activity in most of the BC-ECG signals. It can be concluded that textile silver CREs are more suitable than those of PEDOT:PSS for obtaining BC-ECG records. These developed textile electrodes bring the use of CREs closer to the clinical environment.
No takes yet. Share an insight, caveat, or question.
Lidón-Roger et al. (2018) studied ECG recording. Textile silver concentric ring electrodes vs. PEDOT:PSS concentric ring electrodes and conventional ECG was evaluated on Electric features and performance in detecting ECG signals. Textile silver concentric ring electrodes provided a more stable response than PEDOT:PSS electrodes and higher spatial resolution than conventional ECG for bipolar concentric ECG recordings.