Key result
Concentric ring electrode CMV1 outperforms standard 12-lead ECG for detecting atrial activity.
Why the study?
Theoretical studies indicated that the dimensions of concentric ring electrodes regulate the depth of sensed electric dipoles in ECG recordings, but this had not been experimentally confirmed.
Observational (n=29)
Answers are grounded in this paper and cite their sources.
Electrode ring dimensions and position affect concentric ECG signals; leaves open clinical adoption pending validation studies.
To overcome the limited spatial resolution of standard 12-lead ECG recordings, concentric ring electrodes (CRE) have been proposed to provide valuable data for the diagnosis of a wide range of cardiac abnormalities, including infarction and arrhythmia. Although theoretical studies indicate that the dimensions of the CRE regulate the depth of the electric dipoles sensed by these electrodes, this has not been experimentally confirmed. The aim of this work was to analyze the influence of CRE dimensions and position of a wireless multi-CRE sensor node on the cardiac signal recorded. For this, four wireless multichannel ECG recording nodes based on flexible multi-ring electrodes were placed at positions CMV1 (position comparable to V1), CMV2, CMV4R and CMV5; each node providing three bipolar concentric ECG signals (BC-ECG). Standard 12-lead ECG and 12 BC-ECG signals were recorded in 29 volunteers. The results revealed that a ring with an outer diameter of 33.5 mm achieves a balance between the ease-of-use and spatial resolution of smaller electrodes and improved detectability and higher amplitudes of signals from larger ring electrodes. Although a standard 12-lead ECG outperforms BC-ECC recordings in detectability of cardiac waves, if the relative amplitude of the wave is also considered, BC-ECG at CMV1 proved superior at picking up atrial activity. In fact, in most of the BC-ECG signals picked up at CMV1, P1 and P2 atrial activity waves were more clearly identified than in simultaneous 12-Lead ECG signals. Likewise, BC-ECG signals revealed higher spatial resolution in detecting anomalous electrical activity in local regions, such as impaired intraventricular driving, or atrioventricular blocks. Finally, the wireless multi-CRE sensor node provides enhanced comfort and handling to both patient and clinician over wired systems.
No takes yet. Share an insight, caveat, or question.
Prats-Boluda et al. (2016) conducted an observational in Cardiac abnormalities (including impaired intraventricular driving and atrioventricular blocks) (n=29). Wireless multi-concentric ring electrode (CRE) sensor node recording bipolar concentric ECG (BC-ECG) vs. Standard 12-lead ECG was evaluated on Detectability, spatial resolution, and relative amplitude of cardiac waves (including atrial activity and local anomalous electrical activity). Concentric ring electrodes with a 33.5 mm outer diameter balanced spatial resolution and signal amplitude, with CMV1 proving superior to 12-lead ECG for atrial activity in 29 volunteers.
Synapse has enriched 10 related papers. Consider them for comparative context: