Key result
A flexible 5-pole concentric ring electrode successfully identified intestinal slow waves, with a 30 mm outer ring bipolar configuration and electrolytic gel yielding the best results.
A novel flexible multiring concentric electrode optimizes the non-invasive identification of intestinal slow waves when used in a bipolar configuration with gel.
Prototype enables simultaneous electrode configuration testing; leaves open clinical translation pending in vivo validation.
Developing new types of optimized electrodes for specific biomedical applications can substantially improve the quality of the sensed signals. Concentric ring electrodes have been shown to provide enhanced spatial resolution to that of conventional disc electrodes. A sensor with different electrode sizes and configurations (monopolar, bipolar, etc.) that provides simultaneous records would be very helpful for studying the best signal-sensing arrangement. A 5-pole electrode with an inner disc and four concentric rings of different sizes was developed and tested on surface intestinal myoelectrical recordings from healthy humans. For good adaptation to a curved body surface, the electrode was screen-printed onto a flexible polyester substrate. To facilitate clinical use, it is self-adhesive, incorporates a single connector and can perform dry or wet (with gel) recordings. The results show it to be a versatile electrode that can evaluate the optimal configuration for the identification of the intestinal slow wave and reject undesired interference. A bipolar concentric record with an outer ring diameter of 30 mm, a foam-free adhesive material, and electrolytic gel gave the best results.
No takes yet. Share an insight, caveat, or question.
Zena-Giménez et al. (2018) studied Healthy. Flexible 5-pole multiring concentric electrode was evaluated on Identification of the intestinal slow wave and rejection of undesired interference. A flexible 5-pole concentric ring electrode successfully identified intestinal slow waves, with a 30 mm outer ring bipolar configuration and electrolytic gel yielding the best results.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: