Key result
The active concentric ring electrode sensor presented significantly less ECG and respiration interference than bipolar recordings for non-invasive detection of intestinal myoelectric signals.
Why the study?
Does an active concentric ring electrode sensor reduce ECG and respiration interference in non-invasive detection of intestinal myoelectric signals compared to bipolar recordings?
Does an active concentric ring electrode sensor reduce ECG and respiration interference in non-invasive detection of intestinal myoelectric signals compared to bipolar recordings?
A novel active concentric ring electrode sensor successfully reduces physiological interference (ECG and respiration) during non-invasive abdominal surface recording of intestinal myoelectric signals.
May improve non-invasive intestinal monitoring; hypothesis-generating pending clinical validation.
Although the surface electroenterogram (EEnG) is a weak signal contaminated by strong physiological interference, such as ECG and respiration, abdominal surface recordings of the EEnG could provide a non-invasive method of studying intestinal activity. The goal of this work was to develop a modular, active, low-cost and easy-to-use sensor to obtain a direct estimation of the Laplacian of the EEnG on the abdominal surface in order to enhance the quality of bipolar surface monitoring of intestinal activity. The sensor is made up of a set of 3 concentric dry Ag/AgCl ring electrodes and a battery-powered signal-conditioning circuit. Each section is etched on a different printed circuit board (PCB) and the sections are joined to each other by surface mount technology connectors. This means the sensing electrodes can be treated independently for purposes of maintenance and replacement and the signal conditioning circuit can be re-used. A total of ten recording sessions were carried out on humans. The results show that the surface recordings of the EEnG obtained by the active sensor present significantly less ECG and respiration interference than those obtained by bipolar recordings. In addition, bioelectrical sources whose frequency fitted with the slow wave component of the EEnG (SW) were identified by parametric spectral analysis in the surface signals picked up by the active sensors.
No takes yet. Share an insight, caveat, or question.
Prats-Boluda et al. (2010) studied Intestinal myoelectric signals (n=10). Active concentric ring electrode sensor vs. Bipolar surface recordings was evaluated on ECG and respiration interference in surface recordings of the EEnG. The active concentric ring electrode sensor presented significantly less ECG and respiration interference than bipolar recordings for non-invasive detection of intestinal myoelectric signals.