Permanent neurological injury is a rare but significant complication of unintended intraneural injection during peripheral nerve blockade (PNB) [1]. Injection pressures > 570 mmHg, combined with intrafascicular needle placement, increase the likelihood of nerve damage in animals, since greater force is required to inject a local anaesthetic intraneurally than perineurally, whereas injection pressures < 200 mmHg are not associated with neurological injury [2, 3]. To avoid high pressures, many anaesthetists rely on their own, or their assistant’s, ‘thumb-feel’ perception of the syringe plunger pressure during injection. This method is unreliable as anaesthetists ‘vary widely in their perception of appropriate force and rate of injection during PNB’ [4]. An in-line manometer (BSmart™; Concert Medical LLC, Norwell, MA, USA) is commercially available, alerting the user if injection pressures exceed firstly 775 mmHg, and then 1034 mmHg. This disposable device is available for around £ 6.50 (€ 7.32; US$ 10.44) from the manufacturer’s website [5], but is not marketed in the UK at present. We suggest an alternative, easy and low-cost setup for monitoring injection pressures during PNB. It consists of a 20-ml syringe, connector tubing, three-way tap, 0.9% saline, and an intravenous syringe pump with pressure monitoring. The syringe is filled with 0.9% saline and connected to the block needle tubing via the three-way tap and the connector tubing (Fig. 11). The intravenous syringe pump is set to a minimal infusion rate of 0.1 ml.h−1. The pump display is then selected to show pressure monitoring before injection for PNB is commenced. The set-up for assessing injection pressures during peripheral nerve block; a high injection pressure of 810 mmHg is measured by the syringe pump, alerting to potential intraneural needle tip placement (insert). We use an Injectomat MC Agilia® pump (Fresenius Kabi AG, Bad Homburg, Germany), but any similar device allowing pressure monitoring across an appropriate range should be suitable. Equally, a disposable transducer set for intra-arterial blood or central venous pressure measurement could be used, but at much higher cost. The suggested set-up provides real-time measurement of injection pressures at the syringe port and allows both the anaesthetist performing the PNB and the assistant injecting to assess pressures simultaneously in addition to, and without sacrificing, the traditional ‘thumb-feel’ technique.
No takes yet. Share an insight, caveat, or question.
Siegmueller et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: