Answers are grounded in this paper and cite their sources.
AIM: To evaluate the precision of transfer of a computer-based three-dimensional (3D) planning, using re-formatted cone-beam images, for oral implant placement in partially edentulous jaws. MATERIAL AND METHODS: Four formalin-fixed cadaver jaws were imaged in a 3D Accuitomo FPD cone-beam computed tomography (CT). Data were used to produce an accurate implant planning with a transfer to surgery by means of stereolithographic drill guides. Pre-operative cone-beam CT images were subsequently matched with post-operative ones to calculate the deviation between planned and installed implants. RESULTS: Placed implants (length: 10-15 mm) showed an average angular deviation of 2 degrees (SD: 0.8, range: 0.7-4.0 degrees ) as compared with the planning, while the mean linear deviation was 1.1 mm (SD: 0.7 mm, range 0.3-2.3 mm) at the hex and 2.0 mm (SD: 0.7 mm, range 0.7-2.4 mm) at the tip. CONCLUSIONS: Cone-beam images could be used for implant planning, taking into account a maximal 4 degrees angular and 2.4 mm linear deviation at the apical tip.
No takes yet. Share an insight, caveat, or question.
Assche et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: