This is the authors' abstract. We don't add key points for this paper.
Local gradient coils can improve the performance of echo-planar, diffusion, and short TE imaging in the brain. A modified singular value decomposition algorithm, which allows the rapid optimization and comparison of designs, was employed to optimize head size gradient coils. A theoretical expression for the torque on a cylindrical coil is presented and used to design coils that are free from torque while pulsed within a magnetic field. Gradient coils of various lengths both with and without torque constraints were compared; although torque-free coils do not perform as well as unbalanced coils, asymmetric torque-balanced coils can achieve comparable homogeneity with only a modest increase in inductance and resistance. Both types of coils outperform body size gradient coils by a dramatic margin. A three-axis head gradient designed using these techniques was constructed and used for brain imaging on a clinical scanner.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Alsop et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: