Answers are grounded in this paper and cite their sources.
An average, variable precision error of up to 6% related to source age was observed for dual-photon absorptiometry of the spine in a longitudinal study of bone mineral content involving 393 women. Application of a software correction for source decay compensated for only a portion of this error. The authors conclude that measurement of bone-loss rates using serial dual-photon bone mineral measurements must be interpreted with caution.
No takes yet. Share an insight, caveat, or question.
Ross et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: