Answers are grounded in this paper and cite their sources.
Numerical simulation demonstrates phase-dependent photometric accuracy in FT-IR spectra, highlighting the impact of point selection and apodization.
Two approaches to phase correction in FT-IR spectroscopy are discussed. Using synthetic data, both the Mertz method and the symmetrization-convolution method are evaluated with respect to the accurate reproduction of a single-beam spectrum. The variation in photometric accuracy with (1) the number of points used in the phase calculation, and (2) apodization in the phase calculation is examined.
No takes yet. Share an insight, caveat, or question.
D. Bruce Chase (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: