Key result
Continuous BP monitoring achieves successful measurement in ~69% of geriatric rehabilitation inpatients to detect orthostatic hypotension.
Why the study?
Initial orthostatic hypotension is prevalent in older adults, but continuous blood pressure devices needed for its diagnosis are not widely used in geriatric rehabilitation.
Is continuous beat-to-beat blood pressure monitoring (Finapres) feasible for diagnosing initial orthostatic hypotension in geriatric rehabilitation inpatients?
Cohort (n=29)
No
Is continuous beat-to-beat blood pressure monitoring (Finapres) feasible for diagnosing initial orthostatic hypotension in geriatric rehabilitation inpatients?
Nearly one-third of geriatric rehabilitation inpatients had unsuccessful continuous beat-to-beat blood pressure measurements despite hand warming interventions, highlighting significant practical barriers to routine screening for initial orthostatic hypotension in hospitalized older adults.
Answers are grounded in this paper and cite their sources.
Finapres monitoring faces substantial feasibility barriers in geriatric inpatients; leaves open its role for initial orthostatic hypotension diagnosis pending protocol refinements.
BACKGROUND: Initial orthostatic hypotension (IOH) is highly prevalent in older adults and may interfere with the ability to regain function after acute hospitalization. IOH assessment requires a non-invasive, beat-to-beat continuous blood pressure device, which is not widely used in geriatric rehabilitation. Our aim was to test the feasibility of diagnosing IOH using a continuous blood pressure device in geriatric rehabilitation inpatients. METHODS: Geriatric rehabilitation inpatients of the REStORing Health of Acutely Unwell AdulTs (RESORT) cohort admitted to a tertiary hospital were randomly selected to undergo continuous blood pressure monitoring (Finapres) for 5 min in the supine position and 3 min of standing or sitting when unable to stand. Interventions to warm hands and adjusting the cuff pressure sizes were attempted if no signal was obtained or an error message occurred. RESULTS: Of 37 randomly selected inpatients, 29 {55.2% female; mean age 82.8 (standard deviation [SD]) 6.6 years} agreed to the continuous blood pressure measurement. Successful measurements were achieved in 20 out of 29 inpatients, two after hand warming. Patients with unsuccessful measurements were likely to be older (mean age 87.2 [SD] 4.4 years, p = 0.03), have cerebrovascular disease (p = 0.006), lower body mass index (p = 0.012), and a lower short physical performance battery score (p = 0.039). Eight out of 20 patients had IOH. CONCLUSION: The number of unsuccessful continuous blood pressure measurements was high in a population with high IOH prevalence despite multiple interventions to establish a signal. Future research should focus on improving the efficiency of continuous blood pressure devices in hospitalized patients with unsuccessful signals.
No takes yet. Share an insight, caveat, or question.
Tran et al. (2022) conducted a cohort in Initial orthostatic hypotension in geriatric rehabilitation inpatients (n=29). Continuous non-invasive beat-to-beat blood pressure monitoring (Finapres NOVA) with hand warming and cuff adjustment was evaluated on Feasibility of diagnosing initial orthostatic hypotension (successful continuous blood pressure measurement defined as a signal with measurable waveforms throughout the last 60 s before postural change and 180 s of active stand or sit). Continuous blood pressure monitoring (Finapres) achieved successful measurements in 20 of 29 geriatric rehabilitation inpatients, diagnosing initial orthostatic hypotension in 8 of 20.
Synapse has enriched 10 related papers. Consider them for comparative context: