Key result
Endovascular stenting for iliac dissection restores ECMO flow from 2.3 to 4.2 lpm.
Why the study?
Can endovascular stenting restore extracorporeal membrane oxygenation flow in a patient with common iliac artery dissection complicating peripheral venoarterial ECMO?
Case Report (n=1)
No
Can endovascular stenting restore extracorporeal membrane oxygenation flow in a patient with common iliac artery dissection complicating peripheral venoarterial ECMO?
Endovascular stenting via trans-ECMO or transbrachial access can serve as a rescue strategy to restore peripheral VA-ECMO flow obstructed by iliac artery dissection, though careful stent sizing is critical to prevent kinking.
Answers are grounded in this paper and cite their sources.
May rescue VA-ECMO flow in iliac dissection; leaves open its role as standard rescue pending prospective data.
Ventricular septal defect (VSD) is a known complication after myocardial infarction associated with high mortality. Extracorporeal membrane oxygenation (ECMO) is being successfully used in patients with VSD as a bridge to definitive surgical repair. Although often the only possibility to stabilize hemodynamics and oxygenation, ECMO has many potential complications, carrying significant morbidity and mortality. Here, the patient presented with a postinfarct VSD on peripheral venoarterial ECMO who developed a dissection of the common iliac artery (CIA) on the 5 th day after ECMO implantation. As a result, a sudden drop in ECMO flow has become evident along with high pressures in the arterial cannula. After a definitive diagnosis of a CIA lesion obstructing the blood flow was made, trans-ECMO endovascular repair of CIA was performed. Four days after endovascular repair, we encountered the same problem of decreased blood flow associated with stent kinking and were approached with another endovascular repair to re-establishing full ECMO flow.
No takes yet. Share an insight, caveat, or question.
Depta et al. (2023) conducted a case report in Post-infarction ventricular septal defect on peripheral venoarterial ECMO complicated by common iliac artery dissection (n=1). Trans-ECMO and transbrachial endovascular stenting of the left common iliac artery was evaluated on Restoration of ECMO blood flow and normalization of arterial cannula pressure. Endovascular stenting of a dissected left common iliac artery restored ECMO flow from 2.3 to 4.2 lpm and reduced arterial cannula pressure from 360 to 240 mmHg.