Answers are grounded in this paper and cite their sources.
We have constructed a low-cost, portable, high-sensitivity 16-channel optically pumped magnetometer (OPM), operating in the spin-exchange relaxation-free regime, and demonstrated its applications in magnetocardiography (MCG). The decrease in the cost of sensors by an order of magnitude is achieved by the 16-channel operation realized in a single module using a single large flat pancake rubidium vapor cell, broad pump and probe laser beams, and a 16-channel photodiode array. The OPM is also based on a fiber-coupled nearly parallel-beam configuration to facilitate multichannel design and make the system portable. For human MCG experiments, the 16-channel OPM which includes optical components is placed inside a wooden enclosure for laser safety. The enclosure is set on a nonmagnetic table inside a magnetically shielded room, while the lasers and electronics are placed outside the shielding room, to avoid magnetic noise. We show that the 16-channel OPM enables simultaneous imaging of human cardiac activity on a large area of the chest in a single scan. The multichannel capability will accelerate clinical MCG imaging to reduce the procedure time and patient fatigue.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: