The radio frequency (rf) magnetic field in a microwave cavity ought to be uniform along a line sample in electron paramagnetic resonance (EPR) spectroscopy so that all portions respond uniformly. Mett, Froncisz, and Hyde discovered a way to achieve this objective [Rev. Sci. Instrum. 72, 4188 (2001)]. Their resonators consisted of three regions, a central section for the sample with dimensions at the cutoff condition, and two end sections that had the same cross section as the central section but were made electrically larger by filling them with a low-loss dielectric. The end sections were each one-quarter wavelength long. We have found that the dielectric in the end sections can be omitted and the dimensions made correspondingly larger. Effects of the resulting discontinuities in cavity cross sections perpendicular to the cavity axis have been analyzed using finite element high frequency structure simulator calculations. Closed form expressions for Q values and relative rf field values have been obtained. The length of the uniform field region is decreased somewhat by the discontinuities (∼1 cm at X band). This disadvantage is outweighed by the benefits of higher Q values and elimination of impurity EPR signals from the dielectric materials. End sections may be cylindrical or hemispherical for cylindrical modes and rectangular or hemicircular for rectangular modes.