This is the authors' abstract. We don't add key points for this paper.
We have designed and constructed a quiet gradient set with restricted access for the combined purposes of evaluating the principles of active acoustic screening, recently introduced by Mansfield, Glover, and Bowtell, and for EPI studies of the head at 3.0 T. The design utilizes the return paths of the conductors in a closed arc loop arrangement to eliminate net Lorentz forces thereby attenuating acoustic noise especially at low frequency. This design should significantly reduce the dangers to patients of high noise levels, especially in high field magnetic resonance imaging systems.
No takes yet. Share an insight, caveat, or question.
Mansfield et al. (1995) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: