A new signal processing method for an acoustic sensor in an air pillow improves the accuracy of noninvasive, unconstrained respiration and heartbeat monitoring, especially during weak respiration.
Answers are grounded in this paper and cite their sources.
May enhance unconstrained respiration monitoring accuracy; leaves open clinical validation before practice adoption.
The authors previously developed an unconstrained and noninvasive measurement system of heartbeat and respiration using an acoustic sensor enclosed in an air pillow or an air mat. The system measured instantaneous heartbeat and respiration periods by applying Kalman filters and maximum likelihood method respectively to the dynamic models of the heartbeat and respiration signals extracted from the sensor output.To improve the respiration measurement accuracy especially in a weak respiration case, the paper proposes a measurement system taking into account not only the fundamental mode of the heartbeat, but also the second harmonic component of the heartbeat and the two side band components of the heartbeat made by the respiration.
No takes yet. Share an insight, caveat, or question.
Matsubara et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: