Fatal atrial dislodgement of a Cross-Jones mitral valve occluder due to material wear prompts a recommendation for elective replacement after five years of function.
Fatal atrial dislodgement of a lenticular disc occurred seven years after surgery in a 54-year-old patient, who had had a mitral valve replacement with a Cross-Jones prosthesis, for ruptured chordae tendinae. A marked distortion of the titanium ring reinforced silicone rubber lens disc due to material wear was the cause of this complication. From the literature available to us, atrial dislodgement of a prosthetic mitral occluder has not been previously recorded. We therefore intend to recommend elective replacement of the Cross-Jones prosthesis in all patients who have had their artificial valve functioning for more than five years.
No takes yet. Share an insight, caveat, or question.
Schachner et al. (1979) studied this question.