Key result
A time-domain heart rate measurement algorithm using PPG and accelerometer data achieved comparable result quality to existing algorithms while requiring lower hardware resources.
Why the study?
Does the proposed PPG time-domain algorithm accurately measure heart rate with lower hardware resources compared to existing algorithms?
Does the proposed PPG time-domain algorithm accurately measure heart rate with lower hardware resources compared to existing algorithms?
The proposed PPG time-domain algorithm achieves comparable heart rate measurement accuracy to existing algorithms while requiring fewer hardware resources, making it suitable for resource-constrained wearable devices.
Answers are grounded in this paper and cite their sources.
Real-time PPG heart rate algorithm with accelerometer motion detection may support wearables; leaves open clinical validation before practice use.
This paper presents an algorithm for the measurement of the human heart rate, using photoplethysmography (PPG), i.e., the detection of the light at the skin surface. The signal from the PPG sensor is processed in time-domain; the peaks in the preprocessed and conditioned PPG waveform are detected by using a peak detection algorithm to find the heart rate in real time. Apart from the PPG sensor, the accelerometer is also used to detect body movement and to indicate the moments in time, for which the PPG waveform can be unreliable. This paper describes in detail the signal conditioning path and the modified algorithm, and it also gives an example of implementation in a resource-constrained wrist-wearable device. The algorithm was evaluated by using the publicly available PPG-DaLia dataset containing samples collected during real-life activities with a PPG sensor and accelerometer and with an ECG signal as ground truth. The quality of the results is comparable to the other algorithms from the literature, while the required hardware resources are lower, which can be significant for wearable applications.
No takes yet. Share an insight, caveat, or question.
Marek et al. (2020) studied this question. Photoplethysmographic time-domain heart rate measurement algorithm vs. Other algorithms from the literature was evaluated on Heart rate measurement quality and hardware resource requirements. A time-domain heart rate measurement algorithm using PPG and accelerometer data achieved comparable result quality to existing algorithms while requiring lower hardware resources.