Key result
BOSS balloon-guide catheter achieves ~100% successful reperfusion in stroke thrombectomy.
Observational (n=50)
Non-randomized
Yes
Answers are grounded in this paper and cite their sources.
Observational registry shows high reperfusion and safety with a balloon guide catheter in acute ischemic stroke, indicating strong technical efficacy.
Background We describe the first-in-human experience using the Balloon Occlusion Stroke Sheath (BOSS TM ) balloon-guide catheter to perform stroke thrombectomy in 50 consecutive patients enrolled in the Flow Arrest Safety and Technical success with balloon-guide catheters trial. This aspiration system includes a novel 9.4F balloon-guide catheter conduit for the insertion and guidance of catheters with a balloon providing temporary flow arrest. Methods The Flow Arrest Safety and Technical success with balloon-guide catheter trial is a single-arm, prospective, multi-center, non-randomized, observational registry evaluating the use of the market-released BOSS TM balloon-guide catheter in adult patients diagnosed with an acute ischemic stroke attributable to large vessel occlusion. The purpose of the current trial was to assess the safety and technical success associated with the use of the BOSS TM balloon-guide catheter. Results Fifty patients met inclusion criteria with a mean baseline National Institutes of Health Stroke Scale (NIHSS) of 16. Treatment devices, including aspiration and stent retriever devices, were used in a total of 88 passes. The BOSS TM balloon-guide catheter was compatible with all stroke thrombectomy treatment devices used in 98% (49/50) of procedures. Balloon inflation and flow arrest were achieved in 100% (50/50) and 98% (49/50) of cases, respectively. Balloon deflation and retraction were observed in 100% (50/50) of cases. Successful reperfusion (modified thrombolysis in cerebral infarction score > 2b) was achieved in 100% of cases with single-pass reperfusion achieved in 62% (31/50) of cases. Conclusions The BOSS TM balloon-guide catheter is a safe and technically effective adjunctive device for mechanical thrombectomy of acute ischemic stroke due to large vessel occlusion.
No takes yet. Share an insight, caveat, or question.
Cuoco et al. (2023) conducted an observational in Acute ischemic stroke attributable to large vessel occlusion (n=50). Balloon Occlusion Stroke Sheath (BOSS) 9.4F balloon-guide catheter was evaluated on Safety and technical success (including device compatibility, flow arrest, and successful reperfusion defined as mTICI > 2b). The BOSS balloon-guide catheter achieved 98% flow arrest and 100% successful reperfusion (mTICI >2b), with 62% single-pass reperfusion in stroke thrombectomy.
Synapse has enriched 10 related papers. Consider them for comparative context: