Key result
16-detector row CT coronary angiography (370-ms rotation) achieved optimal image quality across all segments when reconstructed at 60% of the RR interval in patients with heart rates ≤70 bpm.
Observational (n=21)
Answers are grounded in this paper and cite their sources.
OBJECTIVE: To evaluate the best reconstruction window for noninvasive coronary angiography when using a 16-detector row computed tomography (CT) scanner with a gantry rotation time of 370 milliseconds. METHODS: In a pilot study, 189 coronary artery segments of 21 patients with a mean heart rate of 65 beats per minute (bpm, maximum: 45-94 bpm) were investigated using a 16-detector row CT scanner. Raw data were reconstructed in 10% increments from 40% to 70% of the RR interval. Two experienced observers independently evaluated the image quality of the coronary arteries in a segmental fashion. A 5-point ranking scale was applied, with 1 being very poor (no evaluation possible); 2, poor; 3, moderate; 4, good; and 5, very good. RESULTS: In the mean of all patients, the best reconstruction window was found to be at 60% of the RR interval. In patients with higher heart rates, the best reconstruction window was found to be at an earlier stage of the R wave-to-R wave interval. CONCLUSIONS: Initial results show that good diagnostic image quality could be achieved for all evaluated segments of the coronary tree with image reconstructions at 60% of the R wave-to-R wave interval in patients with heart rates of 70 bpm or less. Using a 16-detector row CT scanner with a gantry rotation time of 370 milliseconds, the need for adapting the reconstruction window to each segment for the best image quality was overcome in those cases. In patients with heart rates faster than 70 bpm, reconstructions at an earlier stage within the cardiac cycle were necessary.
No takes yet. Share an insight, caveat, or question.
Bley et al. (2005) conducted an observational in Coronary artery evaluation via computed tomography angiography (n=21). 16-detector row CT coronary angiography with 370-ms gantry rotation reconstructed at 40% to 70% of the RR interval vs. Different reconstruction windows (10% increments from 40% to 70% of the RR interval) was evaluated on Diagnostic image quality of coronary artery segments on a 5-point ranking scale. 16-detector row CT coronary angiography (370-ms rotation) achieved optimal image quality across all segments when reconstructed at 60% of the RR interval in patients with heart rates ≤70 bpm.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: